Evaluation of the clinical sensitivities of three viral load assays with plasma samples from a pediatric population predominantly infected with human immunodeficiency virus type 1 subtype G and BG recombinant forms

Use this identifier to reference this record.
Name:Description:Size:Format: 
03_G_BG_JCM.pdf142.35 KBAdobe PDF Download

Advisor(s)

Abstract(s)

The viral load assays AMPLICOR HIV-1 Monitor Test 1.5, Nuclisens HIV-1 QT, and Quantiplex HIV RNA 3.0 (bDNA) were evaluated for their abilities to quantify human immunodeficiency virus type 1 (HIV-1) RNA in 64 plasma samples from 21 children infected in Portugal. The children were infected with HIV-1 subtypes A1, B, F1, G, and BG recombinant virus. AMPLICOR v1.5 and Quantiplex v3.0 detected all samples, and there was a good correlation of results between the two kits. Thirty-eight specimens containing HIV-1 subtype B, G, or recombinant BG, could not be detected by Nuclisens HIV-1 QT. We also evaluated the new Retina HIV-1 assay on 21 samples that were HIV-1 positive; Retina HIV-1 failed to detect 5 of 11 subtype G specimens. AMPLICOR v1.5 and Quantiplex v3.0 assays may be used for HIV-1 RNA quantification in Portugal, whereas an improvement in sensitivity for subtype G and recombinant BG is required for Nuclisens HIV-1 QT and Retina HIV-1.

Description

Copyright © 2003, American Society for Microbiology. All Rights Reserved.

Keywords

Pedagogical Context

Citation

Antunes R, Figueiredo S, Bártolo I, Pinheiro M, Rosado L, Soares I, et al. Evaluation of the Clinical Sensitivities of Three Viral Load Assays with Plasma Samples from a Pediatric Population Predominantly Infected with Human Immunodeficiency Virus Type 1 Subtype G and BG Recombinant Forms. J Clin Microbiol 2003;41:3361–7. https://doi.org/10.1128/JCM.41.7.3361-3367.2003.

Research Projects

Organizational Units

Journal Issue