Utilize este identificador para referenciar este registo: http://hdl.handle.net/10451/53517
Registo completo
Campo DCValorIdioma
degois.publication.firstPage1787pt_PT
degois.publication.lastPage1790pt_PT
degois.publication.locationMontreal, Canadapt_PT
dc.contributor.authorGodinho, Daniela M.-
dc.contributor.authorFelicio, Joao M.-
dc.contributor.authorFernandes, Carlos A.-
dc.contributor.authorConceicao, Raquel C.-
dc.date.accessioned2022-06-28T10:08:55Z-
dc.date.available2022-06-28T10:08:55Z-
dc.date.issued2020-
dc.identifier.urihttp://hdl.handle.net/10451/53517-
dc.description.abstractMedical Microwave Imaging (MWI) has been studied as a technique to aid breast cancer diagnosis. Several different prototypes have been proposed but most of them require the use of a coupling medium between the antennas and the breast, in order to reduce skin backscattering and avoid refraction effects. The use of dry setups has been addressed and recent publications show promising results. In this paper, we assess the importance of considering refraction effects in the image reconstruction algorithms. To this end, we consider a simplified homogeneous spherical model of the breast and analytically compute the propagating rays through the air-body interface. The comparison of results considering only direct ray propagation or refracted rays shows negligible impact on the accuracy of the images for moderately high permittivity media. Thus, we may avoid the computational burden of calculating the refracted rays in complex shapes.pt_PT
dc.language.isoengpt_PT
dc.relationFundação para a Ciência e a Tecnologia-FCT under the fellowship SFRH/BD/129230/2017pt_PT
dc.relationFundação para a Ciência e a Tecnologia-FCT/MEC (PIDDAC) under the Strategic Programme UIDB/BIO/00645/2020pt_PT
dc.relationFundação para a Ciência e a Tecnologia-FEDER–PT2020 Partnership Agreement under Grant UID/EEA/50008/2019pt_PT
dc.rightsopenAccesspt_PT
dc.subjectBreastpt_PT
dc.subjectMicrowave imagingpt_PT
dc.subjectMicrowave theory and techniquespt_PT
dc.subjectAntennaspt_PT
dc.titleStudy of the Refraction Effects in Microwave Breast Imaging Using a Dry Setuppt_PT
dc.typearticlept_PT
dc.description.versioninfo:eu-repo/semantics/publishedVersionpt_PT
dc.peerreviewedyespt_PT
dc.identifier.doi10.1109/EMBC44109.2020.9176439pt_PT
Aparece nas colecções:IBEB - Artigos em Conferências

Ficheiros deste registo:
Ficheiro Descrição TamanhoFormato 
2020EMBC.pdf446,98 kBAdobe PDFVer/Abrir


FacebookTwitterDeliciousLinkedInDiggGoogle BookmarksMySpace
Formato BibTex MendeleyEndnote 

Todos os registos no repositório estão protegidos por leis de copyright, com todos os direitos reservados.